The Operating Principle of a Precision Open
In VNA SOLT calibration, the Open standard provides a known reference reflection at approximately 0° phase. Unlike a theoretical perfect open (which would reflect 100% of the signal with zero phase shift), a real coaxial open exhibits fringing capacitance at the dielectric-air boundary and a small phase offset due to the physical break in the transmission line. Coaxwave precision opens are characterized so that this fringing capacitance and phase offset are mathematically modeled using a offset phase polynomial — typically a 3rd- or 5th-order function of frequency. The VNA software uses these coefficients to compute the exact expected reflection coefficient at every frequency point, enabling accurate separation of systematic error terms.
Frequency Coverage by Interface:
Interface | Frequency | Primary Calibration Use |
|---|---|---|
N-Type / SMA | DC-18GHz | General-purpose SOLT, sub-6GHz and C-band |
3.5mm | DC-33GHz | Extended K-band SOLT, 5G FR1 |
2.92mm (K-Type) | DC-40GHz | Ka-band SOLT, SatCom uplink |
2.4mm | DC-50GHz | E-band SOLT, 5G FR2 |
1.85mm (V-Type) | DC-67GHz | V-band SOLT, W-band extension |
All Transfer Models (complete open circuit matrix — native & adapter-type):
No. | Model Name | Input Interface | Output (Open Reference) | Freq | Notes |
|---|---|---|---|---|---|
1 | CW-OP50-N-M | N-Type (M) | Open reference plane | DC-18GHz | Native, male |
2 | CW-OP50-N-F | N-Type (F) | Open reference plane | DC-18GHz | Native, female |
3 | CW-OP50-SMA-M | SMA (M) | Open reference plane | DC-18GHz | Native, male |
4 | CW-OP50-SMA-F | SMA (F) | Open reference plane | DC-18GHz | Native, female |
5 | CW-OP50-35-M | 3.5mm (M) | Open reference plane | DC-33GHz | Native, male |
6 | CW-OP50-35-F | 3.5mm (F) | Open reference plane | DC-33GHz | Native, female |
7 | CW-OP50-292-M | 2.92mm (M) | Open reference plane | DC-40GHz | Native, male |
8 | CW-OP50-292-F | 2.92mm (F) | Open reference plane | DC-40GHz | Native, female |
9 | CW-OP50-24-M | 2.4mm (M) | Open reference plane | DC-50GHz | Native, male |
10 | CW-OP50-24-F | 2.4mm (F) | Open reference plane | DC-50GHz | Native, female |
11 | CW-OP50-185-M | 1.85mm (M) | Open reference plane | DC-67GHz | Native, male |
12 | CW-OP50-185-F | 1.85mm (F) | Open reference plane | DC-67GHz | Native, female |
13 | CW-OP50-SMA-35-M | SMA (M) → 3.5mm (F) adapter + open | DC-18GHz | SMA-limited | |
14 | CW-OP50-35-SMA-M | 3.5mm (M) → SMA (F) adapter + open | DC-18GHz | SMA-limited | |
15 | CW-OP50-SMA-292-M | SMA (M) → 2.92mm (F) adapter + open | DC-18GHz | SMA-limited | |
16 | CW-OP50-292-SMA-M | 2.92mm (M) → SMA (F) adapter + open | DC-18GHz | SMA-limited | |
17 | CW-OP50-35-292-M | 3.5mm (M) → 2.92mm (F) adapter + open | DC-33GHz | 3.5mm-limited | |
18 | CW-OP50-292-35-M | 2.92mm (M) → 3.5mm (F) adapter + open | DC-33GHz | 3.5mm-limited | |
19 | CW-OP50-292-24-M | 2.92mm (M) → 2.4mm (F) adapter + open | DC-40GHz | 2.92mm-limited | |
20 | CW-OP50-24-292-M | 2.4mm (M) → 2.92mm (F) adapter + open | DC-40GHz | 2.92mm-limited | |
21 | CW-OP50-24-185-M | 2.4mm (M) → 1.85mm (F) adapter + open | DC-50GHz | 2.4mm-limited | |
22 | CW-OP50-185-24-M | 1.85mm (M) → 2.4mm (F) adapter + open | DC-50GHz | 2.4mm-limited |
💡 Frequency Boundaries for Cross-Interface Models: Always bounded by the lower-rated interface. SMA-based adapters capped at DC-18GHz. 3.5mm-based adapters at DC-33GHz. 2.92mm-based at DC-40GHz. 2.4mm↔1.85mm capped at DC-50GHz (2.4mm limit). For any measurement above the adapter limit, use native open + separate calibration.
Key Performance (independent Coaxwave test system):
Reflection Coefficient: Γ ≈ +1 (magnitude ≥ 0.95, decreasing gradually with frequency due to fringing)
Offset Phase Polynomial: 3rd-order (standard) or 5th-order (metrology) coefficients provided in factory manual
Return Loss (magnitude reference): ≥ 20dB (DC-6GHz); degrades gradually above 6GHz due to fringing capacitance
Impedance: 50Ω reference system
Inspection: 100% VNA full-band S11 magnitude and phase sweep + offset phase polynomial extraction + fringing capacitance characterization. Each unit ships with an independent calibration data manual — measured S11 data, polynomial coefficients, and model curves. Laser-marked with unique serial number.
In SOLT calibration, the Open defines the 0° phase reference against which the VNA measures all other reflections. Coaxwave Precision Opens are not simple "uncovered" connectors — they are calibrated microwave components where the fringing capacitance at the dielectric-air boundary and the physical offset from the reference plane are fully characterized. This enables the VNA to compute directivity and source-match error coefficients with metrology-grade accuracy.
Unlike a fixed short circuit (which has a relatively constant phase near 180°), an open circuit's phase deviates significantly from 0° as frequency increases due to fringing fields at the coaxial aperture. Coaxwave provides offset phase polynomial coefficients (C0, C1, C2, C3, and optionally C4, C5) that model this behavior:
φ(f) = C0 + C1·f + C2·f² + C3·f³ + C4·f⁴ + C5·f⁵
Where φ is the phase offset in degrees and f is frequency in GHz. Entering these coefficients into the VNA calibration kit definition reduces calibration residual error by up to 5× compared to using an "ideal open" model.
Coaxwave provides 12 native models (6 interfaces × M/F) and 10 adapter-type configurations, covering every major cross-interface transition in the RF calibration chain.
Native (interface-to-open, full frequency):
N-Type / SMA: DC-18GHz, M/F
3.5mm: DC-33GHz, M/F
2.92mm: DC-40GHz, M/F
2.4mm: DC-50GHz, M/F
1.85mm: DC-67GHz, M/F
Adapter-Type (via precision transition, frequency bounded by opposite interface):
SMA ↔ 3.5mm: DC-18GHz (SMA-limited)
SMA ↔ 2.92mm: DC-18GHz (SMA-limited)
3.5mm ↔ 2.92mm: DC-33GHz (3.5mm-limited)
2.92mm ↔ 2.4mm: DC-40GHz (2.92mm-limited)
2.4mm ↔ 1.85mm: DC-50GHz (2.4mm-limited)
Every open undergoes three mandatory 100% tests:
VNA Full-Band S11 Magnitude & Phase Sweep: Reflection coefficient measured from DC to the interface's rated frequency. Phase deviation from 0° extracted at every frequency point.
Fringing Capacitance Characterization: Equivalent parallel capacitance at the aperture measured and modeled into the polynomial coefficients.
Mechanical Gauge Verification: Mating depth and center conductor position checked against IEEE 287 / MIL-STD-348 gauge standards.
No sampling, no "typical" curves. Each unit ships with an independent calibration data manual — your unit's actual S11 data, polynomial coefficients, and gauge compliance. Laser-marked serial number guarantees full traceability.
Reflection Coefficient (Γ): ≈ +1 (magnitude ≥ 0.95, decreasing with frequency due to fringing)
Offset Phase Polynomial: 3rd-order standard; 5th-order optional (metrology)
Return Loss: ≥ 20dB reference at low frequencies; degrades gradually above 6GHz per fringing model
Frequency Range:
N-Type / SMA: DC-18GHz
3.5mm: DC-33GHz
2.92mm: DC-40GHz
2.4mm: DC-50GHz
1.85mm: DC-67GHz
Adapter-type: Bounded by lower-rated interface
Connector Interfaces: Per IEEE Std 287™-2007 (1.85/2.4/3.5/2.92mm) and MIL-STD-348 (SMA/N-Type)
Mating Cycles: ≥ 500 (all precision interfaces)
Coupling Torque:
SMA / 3.5mm / N-Type: 0.9 N·m (8 in-lbs)
2.92mm: 0.8–1.1 N·m (7.1–9.7 in-lbs)
2.4mm / 1.85mm: 0.45 N·m (4 in-lbs)
Center Contact: Beryllium copper, gold-plated over nickel underplating
Open Aperture: Precision-machined dielectric break with controlled fringing geometry
Mounting: Inline only (calibration standard, not panel-mount)
Housing / Outer Conductor: Passivated stainless steel (SUS303)
Inner Conductor: Beryllium copper, gold-plated over nickel underplate
Dielectric: PTFE (SMA/N/3.5mm) or PEI (2.92mm/2.4mm/1.85mm) with precision air-gap transition at aperture
Interface Standard: IEEE 287 / MIL-STD-348 per interface
Plating Thickness: Gold ≥ 0.00005" over nickel underplate
Operating Temperature: -55℃ ~ +125℃
Storage Temperature: -65℃ ~ +150℃
Humidity: 95% RH, 240 hours (MIL-STD-202, Method 106)
Vibration: MIL-STD-202, Method 204, Condition D
Shock: MIL-STD-202, Method 213, Condition B
RoHS / REACH: Compliant
Quality System: ISO 9001:2015 / AS9100D
SOLT Calibration Use:
Connect open to VNA port as the second step in SOLT (after Short)
Enter the polynomial coefficients from the factory manual into the VNA calibration kit definition
The VNA uses the coefficients to compute expected S11 at every frequency point
Cross-Interface Use:
When using adapter-type opens, calibrate at the adapter's output interface
Frequency is bounded by the lower-rated side
Handling:
100% VNA phase tested — independent calibration data manual ships with every unit
Keep the open aperture clean and dry — dust or moisture at the aperture changes fringing capacitance → calibration error
Cap when not in use — the open aperture must be protected from contamination
Assembly Environment / System | Recommended Model | Core Advantage | Typical Application |
|---|---|---|---|
5G FR1 ATE | CW-OP50-SMA-M | DC-18GHz, polynomial model | 5G sub-6GHz antenna test |
Aerospace L-band | CW-OP50-N-M | DC-18GHz, rugged | Avionics T/R module calibration |
5G FR2 E-band | CW-OP50-24-M | DC-50GHz, 5th-order poly | 5G mmWave OTA calibration |
SatCom Ka-band | CW-OP50-292-M | DC-40GHz, 3rd-order poly | SatCom HPA calibration |
Automotive Radar (77GHz) | CW-OP50-185-M | DC-67GHz, 5th-order poly | V-band radar module test |
Production ATE | CW-OP50-SMA-F | DC-18GHz, 500+ cycles | High-throughput device test |
Legacy 3.5mm VNA | CW-OP50-35-SMA-M | 3.5mm-to-SMA, DC-18GHz | Extend 3.5mm VNA to SMA DUT |
2.92mm to 2.4mm DUT | CW-OP50-292-24-M | 2.92mm-to-2.4mm, DC-40GHz | Cross-interface SOLT |
Defense EW System | CW-OP50-292-F | DC-40GHz, rugged | EW receiver calibration |
Metrology Lab | CW-OP50-185-M | DC-67GHz, 5th-order poly | National-lab V-band calibration |
Six Interfaces: N-Type, SMA, 3.5mm, 2.92mm, 2.4mm, 1.85mm — M/F each
Full Cross-Interface Adapter Matrix: 10 standard models
Polynomial Order: 3rd-order (standard), 5th-order (metrology) — selectable at order
5th-Order Polynomial (Metrology): For national labs requiring highest phase accuracy above 40GHz
75Ω Version: For broadcast/video and CATV calibration
Phase-Matched Pairs: Two opens matched for differential/multi-channel systems
Panel-Mount Version: For semi-permanent calibration reference planes in ATE fixtures
Captive Center Contact: For high-vibration environments
Keyed Coupling Nut: Prevents mis-mating in confined test bays
Non-Magnetic Version: For EMC chambers and quantum computing RF test
Hermetically Sealed Aperture Cover: Removable protective cap with known electrical effect characterization
Extended Temperature: For thermal chamber calibration
100% VNA Full-Band Phase Test: S11 magnitude and phase from DC to rated frequency
Polynomial Coefficient Data: C0-C5 values in factory calibration manual
Independent Calibration Data Manual: Actual S11 data, coefficients, and model curves
SN Traceability: Laser-marked unique serial number
Rapid Prototyping: Standard models 1-2 week shipment; custom 3-4 weeks
Application Engineering: Coaxwave RF engineers provide:
SOLT calibration kit definition support (polynomial entry guidance)
Cross-interface calibration plane management
Open aperture care and maintenance best practices
Torque and handling best practices per interface
Tip
"There is no perfect standard part, only the most suitable solution."
Standard SOLT Open Step:
Use CW-OP50-[IF]-M (native male open) as the Open standard in a SOLT calibration. Connect to VNA Port A after the Short step. Enter the polynomial coefficients from the factory manual into the VNA's calibration kit definition. The VNA computes the expected reflection at every frequency point using the polynomial model.
Female Open for Reverse Calibration:
Use CW-OP50-[IF]-F when calibrating a test cable or fixture that presents a female connector to the VNA port. This eliminates the need for an additional gender-changer adapter.
SMA VNA to 3.5mm DUT (DC-18GHz):
Use CW-OP50-SMA-35-M. Connect SMA side to VNA. Calibrate at the 3.5mm output plane. Frequency limited to DC-18GHz by SMA.
2.4mm VNA to 1.85mm DUT (DC-50GHz):
Use CW-OP50-24-185-M. Connect 2.4mm side to VNA. Calibrate at 1.85mm plane. Full DC-50GHz coverage. Essential for testing V-band devices on 2.4mm VNAs.
High-Volume Device Test:
Use CW-OP50-SMA-F in automated calibration fixtures. 500+ mating cycles support high-throughput production. The characterized polynomial model ensures consistent calibration across all test sites.
Torque Each Interface to Specification:
SMA/N/3.5mm: 0.9 N·m; 2.92mm: 0.8–1.1 N·m; 2.4mm/1.85mm: 0.45 N·m
Under-tightening causes contact resistance variation → phase error
Over-tightening deforms precision interfaces → reference plane shifts
⚠️ Cross-Interface Frequency Boundaries Are Absolute:
SMA-based adapters capped at DC-18GHz
3.5mm-based adapters capped at DC-33GHz
2.92mm-based adapters capped at DC-40GHz
2.4mm↔1.85mm capped at DC-50GHz
Always Enter Polynomial Coefficients into VNA Software:
Never select "Ideal Open" if you have characterized polynomial data
The coefficients in your factory manual are YOUR unit's actual data — entering them reduces calibration residual error by up to 5×
Protect the Open Aperture:
Dust, oil, or moisture at the aperture changes fringing capacitance → calibration error
Always cap the open when not in use; clean with dry air or lint-free swab if contaminated
Reflection Coefficient: Γ ≈ +1 (magnitude ≥ 0.95, fringing-modeled)
Offset Phase Polynomial: 3rd-order standard; 5th-order metrology optional
Frequency Coverage: DC-18GHz (N/SMA) → DC-33GHz (3.5mm) → DC-40GHz (2.92mm) → DC-50GHz (2.4mm) → DC-67GHz (1.85mm)
Configurations: 12 native models + 10 adapter models = 22 standard transfer models
Materials: Passivated stainless steel + BeCu gold-plated contacts + precision dielectric break
100% VNA Full-Band Phase Test + Polynomial Extraction + Mechanical Gauge + Independent Calibration Data Manual + SN Traceability
Quality System: ISO 9001 / AS9100D | sales@coaxial-rf.com
⚠️ Three Reminders for Open Selection:
An Open is NOT "Nothing Connected": A bare unmated connector is NOT a calibration standard. The fringing capacitance at the dielectric-air boundary must be characterized. Coaxwave opens provide the mathematical model (polynomial coefficients) that lets the VNA account for this effect.
Always Use the Polynomial Data: If your VNA supports offset-delay or polynomial entry for opens, ALWAYS use the Coaxwave-provided coefficients. Using "ideal open" discards the characterization data and increases calibration uncertainty — especially above 10GHz where fringing becomes significant.
The Open Aperture is Delicate: Unlike a short or load, the open has an exposed precision gap. Contamination here directly changes the fringing capacitance. Cap it when not in use. Never touch the center conductor. Clean with dry air only — no solvents at the aperture.
There is no perfect standard part, only the most suitable solution.
Coaxwave Precision Open Circuit Standards are not merely uncovered connectors; they are the engineered 0° phase reference for SOL/SOLT VNA calibration from DC to 67GHz. We machine every passivated stainless-steel body with "German precision" and characterize every unit's fringing model with "French engineering logic". With six interfaces, male/female genders, a complete cross-interface adapter matrix, and 100% VNA certification, Coaxwave delivers the most suitable open standard for your exact SOLT, production ATE, and metrology requirement.
Sales Email: sales@coaxial-rf.com
Company Website: