Coaxwave Precision Short Circuit Standard | Γ=-1 180° | DC-9 to 40GHz | N/SMA/3.5/2.92/2.4mm M-F | Phase ≤±2° | 100% VNA Tested | SOLT VNA Calibration
  • Coaxwave Precision Short Circuit Standard | Γ=-1 180° | DC-9 to 40GHz | N/SMA/3.5/2.92/2.4mm M-F | Phase ≤±2° | 100% VNA Tested | SOLT VNA Calibration
  • 1/1
  • Coaxwave Precision Short Circuit Standard | Γ=-1 180° | DC-9 to 40GHz | N/SMA/3.5/2.92/2.4mm M-F | Phase ≤±2° | 100% VNA Tested | SOLT VNA Calibration

Coaxwave Precision Short Circuit Standard | Γ=-1 180° | DC-9 to 40GHz | N/SMA/3.5/2.92/2.4mm M-F | Phase ≤±2° | 100% VNA Tested | SOLT VNA Calibration

share
WeChat
copy Link
Facebook
X
LinkedIn
Pinterest
Whatsapp
Line
Tumblr
Coaxwave Precision Short Circuit Standard is the core one-port calibration artifact in RF Test & Measurement, engineered for SOLT VNA calibration. 50Ω characteristic impedance, nominal reflection coefficient Γ=-1 (180° phase). Frequency coverage: DC-9GHz (SMA), DC-18GHz (N-Type), DC-26.5GHz (3.5mm), DC-40GHz (2.92mm), DC-40GHz (2.4mm). 100% VNA full-band S-parameter characterized, phase deviation ≤ ±2.0° (DC-6GHz). Full M/F matrix for N/SMA/3.5/2.92/2.4mm. Independent calibration data manual. The definitive reflect standard for correcting reflection tracking, source match, and load match errors in R&D, ATE, aerospace, and metrology.
Get a Quote
DetailTechnical SpecificatAssembly EnvironmentCustomization OptionApplication Scenario

The Operating Principle of a Precision Short

In VNA SOLT calibration, the Short standard provides a known reference reflection. A theoretically perfect short presents a reflection coefficient Γ = -1, meaning the incident signal is reflected with a 180° phase shift and zero magnitude loss. In practice, connector parasitics and the physical location of the shorting plane introduce a small but predictable phase offset from ideal 180°. Coaxwave precision shorts are designed so that this phase offset is tightly controlled and individually characterized. The result: your VNA software can accurately separate systematic error terms, delivering measurement uncertainty limited only by the quality of the calibration standards.

Frequency Coverage by Interface:

Interface

Frequency

Primary Calibration Use

SMA

DC-9GHz / DC-18GHz (selectable)

General-purpose SOLT, sub-6GHz and C-band

N-Type

DC-18GHz

High-power and outdoor SOLT, aerospace

3.5mm

DC-26.5GHz

K-band SOLT, 5G FR1 extended

2.92mm (K-Type)

DC-40GHz

Ka-band SOLT, SatCom uplink

2.4mm

DC-40GHz (interface rated to 50GHz)

E-band SOLT, 5G FR2 (limited by short design to 40GHz)

All Transfer Models (complete short circuit matrix — native & adapter-type):

No.

Model Name

Input Interface

Output (Short Plane)

Freq

Notes

1

CW-SH50-SMA-M

SMA (M)

Short plane internal

DC-9/18GHz

Native, male

2

CW-SH50-SMA-F

SMA (F)

Short plane internal

DC-9/18GHz

Native, female

3

CW-SH50-N-M

N-Type (M)

Short plane internal

DC-18GHz

Native, male

4

CW-SH50-N-F

N-Type (F)

Short plane internal

DC-18GHz

Native, female

5

CW-SH50-35-M

3.5mm (M)

Short plane internal

DC-26.5GHz

Native, male

6

CW-SH50-35-F

3.5mm (F)

Short plane internal

DC-26.5GHz

Native, female

7

CW-SH50-292-M

2.92mm (M)

Short plane internal

DC-40GHz

Native, male

8

CW-SH50-292-F

2.92mm (F)

Short plane internal

DC-40GHz

Native, female

9

CW-SH50-24-M

2.4mm (M)

Short plane internal

DC-40GHz

Native, male

10

CW-SH50-24-F

2.4mm (F)

Short plane internal

DC-40GHz

Native, female

11

CW-SH50-SMA-35-M

SMA (M) → 3.5mm (F) adapter + short

DC-18GHz

SMA-limited


12

CW-SH50-35-SMA-M

3.5mm (M) → SMA (F) adapter + short

DC-18GHz

SMA-limited


13

CW-SH50-SMA-292-M

SMA (M) → 2.92mm (F) adapter + short

DC-18GHz

SMA-limited


14

CW-SH50-292-SMA-M

2.92mm (M) → SMA (F) adapter + short

DC-18GHz

SMA-limited


15

CW-SH50-35-292-M

3.5mm (M) → 2.92mm (F) adapter + short

DC-26.5GHz

3.5mm-limited


16

CW-SH50-292-35-M

2.92mm (M) → 3.5mm (F) adapter + short

DC-26.5GHz

3.5mm-limited


17

CW-SH50-292-24-M

2.92mm (M) → 2.4mm (F) adapter + short

DC-40GHz

2.92mm-limited


18

CW-SH50-24-292-M

2.4mm (M) → 2.92mm (F) adapter + short

DC-40GHz

2.92mm-limited


💡 Frequency Boundaries for Cross-Interface Models: Always bounded by the lower-rated interface. SMA-based adapters capped at DC-18GHz. 3.5mm-based at DC-26.5GHz. 2.92mm↔2.4mm capped at DC-40GHz (2.92mm limit). For any measurement above the adapter limit, use native short + separate calibration.

Key Performance (independent Coaxwave test system):

  • Reflection Coefficient: Γ ≈ -1 (magnitude ≥ 0.98)

  • Phase Deviation from 180°: ≤ ±2.0° (DC-6GHz); ≤ ±3.0° (DC-18GHz); ≤ ±5.0° (DC-40GHz)

  • Return Loss: ≥ 30dB (DC-6GHz); ≥ 25dB (DC-18GHz); ≥ 20dB (DC-40GHz)

  • Impedance: 50Ω reference system

Inspection: 100% VNA full-band S11 phase sweep + short-plane continuity verification + phase deviation extraction. Each unit ships with an independent calibration data manual — measured S11 magnitude/phase, extracted phase offset, and model data. Laser-marked with unique serial number.

🎯 Core Positioning: The Phase-Reference for SOLT Reflection Accuracy

In SOLT calibration, the Short defines the 180° phase reference against which the VNA measures all other reflections. Coaxwave Precision Shorts are not simple "zero-ohm" connectors — they are calibrated microwave components where the physical location of the shorting plane and its parasitic reactance are fully characterized. This enables the VNA to compute directivity and source-match error coefficients with metrology-grade accuracy.

⚡ Phase Accuracy as the Differentiator

  • DC-6GHz: Phase deviation ≤ ±2.0° — suitable for sub-6GHz 5G, GNSS, and CATV calibration where phase accuracy directly impacts return-loss uncertainty.

  • DC-18GHz: Phase deviation ≤ ±3.0° — covers all C/X/Ku-band applications.

  • DC-40GHz: Phase deviation ≤ ±5.0° — Ka-band SOLT with confidence.

🔧 Full Native & Adapter-Type Matrix

Coaxwave provides 10 native models (5 interfaces × M/F) and 8 adapter-type configurations, covering every major cross-interface transition in the RF calibration chain.

Native (interface-to-short, full frequency):

  • SMA: DC-9GHz or DC-18GHz, M/F

  • N-Type: DC-18GHz, M/F

  • 3.5mm: DC-26.5GHz, M/F

  • 2.92mm: DC-40GHz, M/F

  • 2.4mm: DC-40GHz, M/F

Adapter-Type (via precision transition, frequency bounded by opposite interface):

  • SMA ↔ 3.5mm: DC-18GHz (SMA-limited)

  • SMA ↔ 2.92mm: DC-18GHz (SMA-limited)

  • 3.5mm ↔ 2.92mm: DC-26.5GHz (3.5mm-limited)

  • 2.92mm ↔ 2.4mm: DC-40GHz (2.92mm-limited)

🛡️ Coaxwave Quality Barrier

Every short undergoes three mandatory 100% tests:

  1. VNA Full-Band S11 Phase Sweep: Reflection coefficient magnitude and phase measured from DC to the interface's rated frequency. Phase deviation from 180° extracted at every frequency point.

  2. Short-Plane Continuity Check: DC resistance of the shorting element verified ≤ 10 mΩ using 4-wire Kelvin method.

  3. Mechanical Gauge Verification: Mating depth and center conductor position checked against IEEE 287 / MIL-STD-348 gauge standards.

No sampling, no "typical" curves. Each unit ships with an independent calibration data manual — your unit's actual S11 data, phase deviation table, and gauge compliance. Laser-marked serial number guarantees full traceability.


Electrical Performance

  • Reflection Coefficient (Γ): ≈ -1 (magnitude ≥ 0.98)

  • Phase Deviation from 180°: ≤ ±2.0° (DC-6GHz); ≤ ±3.0° (DC-18GHz); ≤ ±5.0° (DC-40GHz)

  • Return Loss: ≥ 30dB (DC-6GHz); ≥ 25dB (DC-18GHz); ≥ 20dB (DC-40GHz)

  • Frequency Range:

    • SMA: DC-9GHz / DC-18GHz (selectable at order)

    • N-Type: DC-18GHz

    • 3.5mm: DC-26.5GHz

    • 2.92mm: DC-40GHz

    • 2.4mm: DC-40GHz (interface rated to 50GHz, short standard optimized to 40GHz)

    • Adapter-type: Bounded by lower-rated interface


Mechanical Performance

  • Connector Interfaces: Per IEEE Std 287™-2007 (3.5/2.92/2.4mm) and MIL-STD-348 (SMA/N-Type)

  • Mating Cycles: ≥ 500 (all precision interfaces)

  • Coupling Torque:

    • SMA / 3.5mm / N-Type: 0.9 N·m (8 in-lbs)

    • 2.92mm: 0.8–1.1 N·m (7.1–9.7 in-lbs)

    • 2.4mm: 0.45 N·m (4 in-lbs)


  • Center Contact: Beryllium copper, gold-plated over nickel underplating

  • Short-Circuit Plane: Gold-plated BeCu button, ≤ 0.005mm flatness, soldered to inner conductor

  • Mounting: Inline only (calibration standard, not panel-mount)

Materials & Structure

  • Housing / Outer Conductor: Passivated stainless steel (SUS303)

  • Inner Conductor: Beryllium copper, gold-plated over nickel underplate

  • Dielectric: PTFE (SMA/N/3.5mm) or air-dielectric (2.92mm/2.4mm at mating plane)

  • Short Element: BeCu disc, gold-plated, ≤ 0.01mm thickness

  • Interface Standard: IEEE 287 / MIL-STD-348 per interface

  • Plating Thickness: Gold ≥ 0.00005" over nickel underplate

Environment & Compliance

  • Operating Temperature: -55℃ ~ +125℃

  • Storage Temperature: -65℃ ~ +150℃

  • Humidity: 95% RH, 240 hours (MIL-STD-202, Method 106)

  • Vibration: MIL-STD-202, Method 204, Condition D

  • Shock: MIL-STD-202, Method 213, Condition B

  • RoHS / REACH: Compliant

  • Quality System: ISO 9001:2015 / AS9100D

Adaptation & Installation

  • SOLT Calibration Use:

    • Connect short to VNA port as the first step in SOLT

    • VNA measures S11 — the 180° phase reference establishes directivity

    • Enter the phase offset data from the factory manual into the calibration kit definition for highest accuracy


  • Cross-Interface Use:

    • When using adapter-type shorts, calibrate at the adapter's output interface

    • Frequency is bounded by the lower-rated side


  • Handling:

    • 100% VNA phase tested — independent calibration data manual ships with every unit

    • Clean mating surfaces with lint-free swabs and IPA before every use

    • Do not over-tighten — excess torque deforms the outer conductor and shifts the short-plane reference



Assembly Environment / System

Recommended Model

Core Advantage

Typical Application

Sub-6GHz 5G ATE

CW-SH50-SMA-M

DC-9GHz, phase ≤±2°

5G FR1 antenna return-loss test

Aerospace L-band Test

CW-SH50-N-M

DC-18GHz, rugged

Avionics T/R module calibration

SatCom Ka-band SOLT

CW-SH50-292-M

DC-40GHz, phase ≤±5°

SatCom HPA output calibration

Automotive Radar (35GHz)

CW-SH50-292-M

DC-40GHz, native

Radar module reflectometry

Production ATE (High Volume)

CW-SH50-SMA-F

DC-18GHz, 500+ cycles

High-throughput device test

Legacy 3.5mm VNA

CW-SH50-35-SMA-M

3.5mm-to-SMA, DC-18GHz

Extend 3.5mm VNA to SMA DUT

2.92mm to 2.4mm DUT

CW-SH50-292-24-M

2.92mm-to-2.4mm, DC-40GHz

Cross-interface SOLT

Defense EW Receiver

CW-SH50-292-F

DC-40GHz, rugged

EW system port calibration

Metrology Lab (18GHz)

CW-SH50-SMA-M

DC-18GHz, characterized

Primary calibration standard

Field Service Kit

CW-SH50-N-F

DC-18GHz, weatherproof

On-site system calibration

Standard Matrix & Extensions

  • Five Interfaces: SMA (DC-9/18GHz), N-Type, 3.5mm, 2.92mm, 2.4mm — M/F each

  • Full Cross-Interface Adapter Matrix: 8 standard models

  • Offset Short (Delay Short): Custom offset lengths (e.g., 1.4mm, 2.5mm, 3.8mm) for LRM/LRL calibration — consult Coaxwave engineering

⚡ Electrical Performance Customization

  • Ultra-Low Phase Deviation (Metrology): Target ≤ ±1.0° (DC-6GHz) for national-lab calibration

  • 75Ω Version: For broadcast/video and CATV calibration

  • Phase-Matched Pairs: Two shorts matched to ±0.5° for differential/multi-channel systems

Structural Customization

  • Panel-Mount Version: For semi-permanent calibration reference planes in ATE fixtures

  • Captive Center Contact: For high-vibration ATE environments

  • Keyed Coupling Nut: Prevents mis-mating in confined test bays

️ Material & Environmental Customization

  • Non-Magnetic Version: For EMC chambers and quantum computing RF test

  • Hermetically Sealed: For pressurized calibration environments

  • Extended Temperature: For thermal chamber calibration (-55℃ to +125℃ operational)

Service & Support

  • 100% VNA Full-Band Phase Test: S11 magnitude and phase from DC to rated frequency

  • Phase Deviation Data: Tabulated phase offset in factory calibration manual

  • Independent Calibration Data Manual: Actual S11 data, phase table, gauge compliance

  • SN Traceability: Laser-marked unique serial number

  • Rapid Prototyping: Standard models 1-2 week shipment; custom 3-4 weeks

  • Application Engineering: Coaxwave RF engineers provide:

    • SOLT calibration kit definition support

    • Phase offset data integration into VNA software

    • Cross-interface calibration plane management

    • Torque and handling best practices per interface


Tip

"There is no perfect standard part, only the most suitable solution."


SOLT Calibration (Same-Interface)

  • Standard SOLT Short Step:
    Use CW-SH50-[IF]-M (native male short) as the Short standard in a SOLT calibration. Connect to VNA Port A. The VNA measures S11 — the 180° phase reference against which Open and Load are compared. For highest accuracy, enter the characterized phase offset from the factory manual into the VNA's calibration kit definition.

  • Female Short for Reverse Calibration:
    Use CW-SH50-[IF]-F when calibrating a test cable or fixture that presents a female connector to the VNA port. This eliminates the need for an additional gender-changer adapter.

Cross-Interface Calibration

  • SMA VNA to 3.5mm DUT (DC-18GHz):
    Use CW-SH50-SMA-35-M. Connect SMA side to VNA. Calibrate at the 3.5mm output plane. Frequency limited to DC-18GHz by SMA.

  • 2.92mm VNA to 2.4mm DUT (DC-40GHz):
    Use CW-SH50-292-24-M. Connect 2.92mm side to VNA. Calibrate at 2.4mm plane. Full DC-40GHz coverage. Essential for testing E-band devices on 2.92mm VNAs.

Production ATE

  • High-Volume Device Test:
    Use CW-SH50-SMA-F in automated calibration fixtures. 500+ mating cycles support high-throughput production. The characterized phase offset ensures consistent calibration across all test sites.

⚠️ Usage Red Lines

  • Torque Each Interface to Specification:

    • SMA/N/3.5mm: 0.9 N·m; 2.92mm: 0.8–1.1 N·m; 2.4mm: 0.45 N·m

    • Under-tightening causes contact resistance variation → phase error between calibrations

    • Over-tightening deforms precision interfaces → the short-plane reference shifts → calibration failure


  • ⚠️ Cross-Interface Frequency Boundaries Are Absolute:

    • SMA-based adapters capped at DC-18GHz — even if the other side is 2.92mm or 3.5mm

    • 3.5mm-based adapters capped at DC-26.5GHz

    • For any measurement above these limits, use native short + separate calibration


  • Always Enter Phase Offset Data into VNA Software:

    • Never assume "ideal short" in calibration kit definition

    • The characterized phase offset in your factory manual is YOUR unit's actual data — entering it reduces calibration residual error by up to 3×


  • 100% VNA Phase Tested — Trust Your Data:

    • Every Coaxwave short ships with S11 phase data across the full frequency range

    • At 40GHz, a 2° phase error translates to significant reflection-tracking uncertainty



Coaxwave Precision Short Circuit Standard · Delivery Commitment

  • Reflection Coefficient: Γ ≈ -1 (magnitude ≥ 0.98)

  • Phase Deviation: ≤ ±2.0° (DC-6GHz) / ≤ ±3.0° (DC-18GHz) / ≤ ±5.0° (DC-40GHz)

  • Frequency Coverage: DC-9/18GHz (SMA) → DC-18GHz (N) → DC-26.5GHz (3.5mm) → DC-40GHz (2.92/2.4mm)

  • Configurations: 10 native models + 8 adapter models = 18 standard transfer models

  • Materials: Passivated stainless steel + BeCu gold-plated contacts + precision short-plane BeCu button

  • 100% VNA Full-Band Phase Test + Short-Plane Continuity + Mechanical Gauge + Independent Calibration Data Manual + SN Traceability

  • Quality System: ISO 9001 / AS9100D | sales@coaxial-rf.com

⚠️ Three Reminders for Short Selection:

  1. A Short is Not a "Zero-Ohm" Adapter: It is a calibrated microwave component with a precisely known phase offset from ideal 180°. Always enter the phase offset from your factory manual into the VNA calibration kit definition — never select "Ideal Short" if you have characterized data.

  2. SMA DC-9GHz vs. DC-18GHz: For general-purpose work, DC-18GHz covers everything. For cost-sensitive sub-6GHz ATE, the DC-9GHz version provides identical phase accuracy within its band. Specifying DC-18GHz for a 2GHz application is acceptable; specifying DC-9GHz for an 18GHz VNA calibration is a measurement error.

  3. 2.4mm Shorts Are Rated to 40GHz: While the 2.4mm interface itself extends to 50GHz, the shorting structure is optimized for the 40GHz calibration band. For measurements above 40GHz on 2.4mm, use a sliding load (SOLR) or offset short for superior phase diversity.


There is no perfect standard part, only the most suitable solution.

Coaxwave Precision Short Circuit Standards are not merely brass buttons in a coaxial shell; they are the engineered 180° phase reference for SOLT VNA calibration from DC to 40GHz. We machine every passivated stainless-steel body with "German precision" and characterize every unit's phase deviation with "French engineering logic". With five interfaces, male/female genders, a complete cross-interface adapter matrix, and 100% VNA certification, Coaxwave delivers the most suitable short standard for your exact SOLT, production ATE, and metrology requirement.

Sales Email: sales@coaxial-rf.com

Company Website: www.coaxial-rf.com