Coaxwave RF & Microwave Probe Product Matrix: Interface, Configuration, Frequency & Application

From DC to 110GHz — One Matrix Table to Master Probe Selection for 5G, Automotive Radar, and 6G THz

2026-09-06 20:47 coaxwave
2

Introduction — Probe Engineering at the Tip of a Needle

The essence of an RF/microwave probe is adaptation: it translates micro-scale bond pads on a chip into standard coaxial or waveguide ports that a vector network analyzer can recognize. Compared to testing after packaging and bonding, on-wafer testing eliminates parasitic parameters introduced by packages and bond wires, enabling a more accurate reflection of the chip's true RF characteristics .

In the rapidly evolving landscape of 5G, automotive radar, and 6G technologies, probes are no longer consumables selected merely to "get the job done." They are core components that directly determine device modeling accuracy, yield analysis, and R&D efficiency. Drawing upon years of RF interconnect expertise, Coaxwave has built a comprehensive probe product matrix spanning from low frequencies to terahertz.

Why Connector Type Defines the Frequency Ceiling

The mechanical structure of a probe's rear connector directly determines its maximum supported frequency. Coaxwave probes strictly adhere to industry-standard connector frequency boundaries :

Connector Type

Maximum Rated Frequency

Mechanical Compatibility

SMA / 3.5mm

26.5 GHz

Mutually compatible SMA

2.92mm (K-type)

40 GHz

Compatible with SMA/3.5mm (reduced bandwidth)

2.4mm

50 GHz

—

1.85mm (V-type)

67 GHz

—

1.0mm

110 GHz (extendable to 130 GHz)

—

0.8mm

145 GHz

—

💡 Selection tip: Beyond the connector's frequency grade, the actual bandwidth requirement of your application is the true basis for selection. S-parameter testing demands probes rated to fully cover the test range; TDR testing can use probes rated at twice the test frequency; signal integrity verification calls for probes rated 1.5-2× the S-parameter bandwidth .

Coaxwave Probe Product Matrix

The following table summarizes the core parameters of Coaxwave's flagship probe series, covering every scenario from low-frequency PCB debugging to W-band mmWave on-wafer testing.

Series

Frequency

Interface

Typical Config.

Insertion Loss

Application Scene

CW-RF20

DC~20 GHz

3.5mm(F) / SMA

GS / GSG

< 1.0 dB

PCBA RF testing, signal integrity verification, circuit debugging

CW-40A

DC~40 GHz

2.92mm(F) K-type

GS / GSG / GSSG

≤ 0.8 dB

5G RF chips (n77/n78/n79), Sub-6GHz RF front-ends, filter & PA wafer-level S-parameter testing

CW-50A

DC~50 GHz

2.4mm(F)

GS / GSG / GSSG

≤ 1.0 dB

Sub-6GHz to mmWave transition band, broadband RF IC R&D

CW-67A

DC~67 GHz

1.85mm(F) V-type

GS / GSG

≤ 1.1 dB

5G FR2 (28GHz/39GHz), early-stage automotive radar R&D

CW-110H

DC~110 GHz

1.0mm(F)

GS / GSG

≤ 1.5 dB

77GHz automotive radar, 6G frontier R&D (100-300GHz), THz front-end devices, high-speed silicon photonics

CW-WR Series*

60~325 GHz

Waveguide

Waveguide probe

Per model

THz imaging, quantum devices, sub-mmWave scientific research

* The CW-WR series covers WR-22 (33-50GHz) through WR-2.2 (325-500GHz) waveguide bands. Detailed specs available in the selection handbook.

Tip Configuration: How to Choose GS / GSG / GSSG?

The tip configuration (pad configuration) determines what type of device a probe can test :

  • GS (Ground-Signal): Single-ended port testing. Suitable for S-parameter measurements of single-ended RF devices. However, GS or SG setups are more prone to imbalance, which can limit performance at higher frequencies .

  • GSG (Ground-Signal-Ground): The most mainstream on-wafer configuration. The ground-signal-ground structure provides excellent common-mode rejection and is the ideal setup for maintaining signal balance and achieving the highest bandwidth . Suitable for bidirectional S-parameter testing of most RF ICs, antenna switches, and LNAs.

  • GSSG (Ground-Signal-Signal-Ground): Differential configuration, dedicated to differential devices such as differential amplifiers, differential antennas, and high-speed SerDes channels. Note: GSSG is more susceptible to unwanted signal coupling, especially at high frequencies .

  • GSGSG: Best choice when you need to reduce crosstalk between adjacent signals .

Tip pitch is another critical parameter. For measurements up to 40GHz, aim for tip pitches under 250µm; at higher frequencies, smaller pitches are necessary to prevent signal distortion caused by ground-path inductance . Coaxwave probes support tip pitches from 50µm to 1250µm depending on the series, with standard tip width of 30µm and a <25µm narrow-tip option available.

Typical Application Scenarios & Probe Matching

Scenario 1: 5G RF Chip On-Wafer S-Parameter Testing

5G terminal and base station chips span Sub-6GHz bands (n77/n78/n79) and FR2 mmWave bands (n257/n258/n260) .

  • Sub-6GHz chips: CW-40A (2.92mm, GSG config) — insertion loss ≤0.8dB, the workhorse for university research and enterprise R&D.

  • FR2 RF front-end modules: CW-67A (1.85mm) — supports full-parameter testing at 28GHz/39GHz.

Scenario 2: 77GHz Automotive Radar MMIC Testing

Automotive radar chips operating in the 77-81GHz band demand extremely high probe phase stability and tip repeatability .

  • Recommended: CW-110H (1.0mm, GSG config), tip pitch 100µm or 150µm. Nickel-alloy tip variants are available for aluminum pad testing.

  • Key metrics: At 77GHz, insertion loss < 1.5dB, return loss < -14dB, ensuring S-parameter and noise figure measurement accuracy for radar MMICs.

Scenario 3: 6G & Terahertz Frontier R&D

Targeting 6G research at 100-300GHz, terahertz imaging, and high-speed digital device testing .

  • Recommended: CW-WR series waveguide probes — multiple waveguide interfaces including WR-10 (75-110GHz), WR-8 (90-140GHz), WR-6 (110-170GHz).

  • Special options: Bias-Tee-integrated probe variants support on-wafer DC bias overlay RF testing.

Scenario 4: PCBA RF Testing & Signal Integrity

For PCB-level RF test points and coaxial debug points .

  • Recommended: CW-RF20 series, SMA or 3.5mm interface, DC~20GHz frequency coverage. Supports signal integrity verification, chip evaluation, coplanar waveguide testing, and circuit debugging.

  • Configuration: GS / GSG, tip pitch options of 800µm and 1500µm.

Engineering Selection Guidelines

Based on Coaxwave's experience in the semiconductor test industry, we offer the following guidance to on-wafer test engineers:

  1. Allow frequency margin: The probe's rated frequency must fully cover — or exceed — the test range to avoid result distortion caused by bandwidth limitations .

  2. Match tip pitch to pad size: Tip pitches from 50µm to 1250µm are available with standard 30µm tip width and a <25µm narrow-tip option. Precisely match to the chip pad layout to avoid test gold stacking.

  3. Match material to pad metal: Gold pads use standard tungsten tips; aluminum pads must use nickel-alloy tips (e.g., CW-40A variants) — otherwise contact resistance will be unstable.

  4. Do not mix calibration substrates: Probes and calibration substrates from different vendors generally cannot be mixed. Use Coaxwave original calibration substrates for SOLT calibration. For >40GHz mmWave MMIC characterization, TRL (Through-Reflect-Line) calibration using transmission line standards fabricated on the same substrate as the DUT is the preferred method .

  5. Manage repeatability: Probes are precision consumables with finely polished, specially chamfered tips. Establish a tip-life tracking system and return for refurbishment when contact resistance rises or S-parameter repeatability degrades.

Conclusion

From DC to 110GHz, from GS single-ended to GSSG differential, from 5G chips to terahertz front-ends — probe selection is essentially a deep understanding of the test object's physical characteristics. A good product matrix table is not a pile of parameters, but a map that lets engineers find "the one" at a glance.

Coaxwave consistently advocates: There is no perfect standard part, only the most suitable solution. Whether it is a standard CW-series on-wafer probe or a custom probe based on special pad layouts, unique frequency bands, or specific environmental constraints, our engineering team starts at the micron scale of the probe tip to deliver the optimal solution for your test system.


📧 Technical Inquiry: sales@coaxial-rf.com